Measuring breast volume in hypertrophy: laser scanning or water displacement?

Tamara Crittenden BAppSc (Hons),1,2 Daisy E Veitch,3 Maciej Henneberg PhD DS,4 Karen Burford BAppSc OT (Hons),1 Phillipa van Essen BHSc MPH,1 Kaveeta Deut BAppSc OT,1 Kalavani Zeitouneh BAppSc OT (Hons),1 Elizabeth Lomax BAppSc OT,1 Philip Griffin MBBS FRACS (Plas),1 Nicola R Dean PhD FRACS (Plast)1,2

 

1 

Department of Plastic and Reconstructive Surgery
Flinders Medical Centre
Bedford Park, South Australia
AUSTRALIA

 

2 

College of Medicine and Public Health
Flinders University
Bedford Park, South Australia
AUSTRALIA

3

Sharp Dummies Pty Ltd
Belair, South Australia
AUSTRALIA

 

4

Adelaide Medical School
University of Adelaide
Adelaide, South Australia
AUSTRALIA

OPEN ACCESS

Correspondence

Name: Nicola R Dean

Address: Department of Plastic and Reconstructive Surgery
Level 5, Flinders Medical Centre
Flinders Drive
Bedford Park, South Australia, 5042
AUSTRALIA

Email: nicola.dean@sa.gov.au

Telephone: +61 8 82045213

Citation: Crittenden T, Veitch DE, Henneberg M, Burford K, van Essen P, Deut K, Zeitouneh K, Lomax E, Griffin P, Dean NR. Measuring breast volume in hypertrophy: laser scanning or water displacement? Aust J Plast Surg. 2018;1(2):33–40.

Accepted for publication: 23 May 2018

Copyright © 2018. Authors retain their copyright in the article. This is an open access article distributed under the Creative Commons Attribution Licence which permits unrestricted use, distribution and reproduction in any medium, provided the original work is properly cited.


Abstract

Background: The accurate determination of intact breast volume facilitates preoperative planning for a range of plastic surgical breast procedures. In women with breast hypertrophy, volumetric assessment assists in planning the amount of tissue to be removed during breast reduction surgery to achieve breast symmetry. Further, in jurisdictions where restrictive surgical guidelines exist, measurement of intact breast volume is essential in order to justify breast reduction surgery. Not all practitioners have access to magnetic resonance imaging (MRI) or three dimensional (3D) laser scanning facilities, so the purpose of this study was to determine whether water displacement of the intact breast is an effective substitute method of measurement in women with breast hypertrophy.

Methods: A prospective cohort study was undertaken to measure breast volume using water displacement and 3D laser scanning in breast hypertrophy patients. The volume of a total of 322 breasts were determined using both measurement techniques; 194 preoperatively and 128 at 12 months following breast reduction surgery. Pearson correlations, linear regression and Bland-Altman analyses were used to compare the methods of breast volume assessment.

Results: The mean breast volume according to 3D laser scan was 1440 millilitres (SD = 588 millilitres) and for water displacement was 1419 millilitres (SD = 811 millilitres). There was a strong linear association between breast volumes as measured using water displacement and 3D laser scanning using a Pearson correlation (r = 0.89, p < 0.001). However, using the Bland-Altman analysis, the two methods were found not to be in agreement, with water displacement values consistently larger than 3D scan values.

Conclusions: While a strong linear association was found between breast volume measurements made using the water displacement technique and 3D laser scanning, the two methods did not have a high level of equivalence. Water displacement of the intact breast is not an accurate enough tool to be used as a substitute for 3D laser scanning, and therefore not an optimal method of breast volume measurement in women with breast hypertrophy.

Key words: breast volume, water displacement, 3D laser scan, breast hypertrophy, breast reduction


Introduction

Breast hypertrophy is for some women a cause of considerable physical and psychosocial impairment, and adversely affects quality of life. Physical symptoms may include: chronic back, neck, shoulder and breast pain; skin rashes; headaches; shoulder grooving from the constant pressure of bra straps; a reduced capacity for carrying out daily activities and exercise; and numbness or tingling in hands/fingers.1 Psychosocially, women with breast hypertrophy often display low self-esteem and poor body image, reduced psychosexual function, anxiety and depression.2

Breast reduction surgery is widely known to be the most effective treatment for breast hypertrophy.3 Surgery provides almost immediate symptomatic relief in most cases, and considerably improves the health-related quality of life and wellbeing in women suffering from the functional symptoms of breast hypertrophy.4,5 Despite these proven benefits, an increase in the prevalence of breast hypertrophy, and hence in demand for breast reductions, has led to restrictions being placed on surgery by healthcare funders and third-party providers in many jurisdictions. Arbitrary restrictions on either the minimum amount of breast tissue required to be removed at surgery or a body mass index cut-off point have been introduced in many institutions and countries worldwide.6

Measurement of breast volume plays an important role in breast reduction, reconstruction, developmental asymmetry and augmentation. In women with breast hypertrophy it is important for preoperative planning, in intraoperative decision-making regarding the amount of tissue to be taken from each breast to achieve symmetry, and where removal of a minimum amount of breast tissue is required to justify surgery.7-11 Accurate estimation of breast volume and predicted tissue resection weight also promotes improved counsel to the patient and provides a valuable guide to training surgeons.

A variety of techniques are described in the literature to meet the need for accurate and objective measurement of breast volumes in the clinical setting including magnetic resonance imaging (MRI), mammography, plaster casting, Grossman-Roudner plastic cups, water displacement, anthropometric measurement and 3D surface imaging.12-21

In a situation where a woman is undergoing post-mastectomy breast reconstruction, the gold standard for measuring the breast tissue volume for reconstruction is the Archimedes method of water displacement of the mastectomy specimen.22 When measuring breast tissue volume in the intact state, such as in candidates for breast reduction surgery, MRI is reported as being the most accurate.23 However, MRI is expensive and not always accessible. Three-dimensional laser scanning has also been demonstrated as a valid method of breast volume measurement,16,17,22 but some centres lack access to this technology. Water displacement of the intact breast has been used in some centres since 1970, but has not previously been validated against more recent methods.24

The purpose of this study is to assess the validity of measuring breast tissue volume using water displacement of the intact breast and to compare this measurement technique with 3D laser scan measurement.

Methods

Patient selection and measurement

A prospective cohort study was performed at Flinders Medical Centre in Adelaide, Australia. All women aged 18 years and above with breast hypertrophy who were eligible to be placed on a waiting list for bilateral breast reduction surgery between March 2007 and February 2016 were informed of the study. Women who met the eligibility criteria were invited to voluntarily participate in the study and were provided with a letter and information sheet further outlining the study. Women who were unable to complete written questionnaires, had impaired cognitive capacity, or refused to participate, were excluded from the study. Approval was obtained for this study from the local Southern Adelaide Clinical Human Research Ethics Committee (SAC HREC approval 118.056).

Bilateral breast volumes were obtained preoperatively and/or at 12 months postoperatively by two different techniques—3D laser scanning and water displacement. Patients proceeded immediately from measurement by water displacement to the 3D scanning within the same appointment to allow direct comparison. Some participants had breast volume measurements taken at both the preoperative and 12 months postoperative times, while other participants were measured once as additional measurement was considered onerous for some patients.

Three-dimensional laser body scanning was performed using a Cyberware WBX scanner (Cyberware, Monterey, USA) and Cyslice software (Headus Pty Ltd, Perth, Australia). The Cyberware WBX scanner has four sets of laser heads and cameras that move vertically to scan the surface anatomy of subjects in the upright position. Prior to scanning, the assessor placed ‘landmarks’ on subjects using palpation to demarcate selected anatomical areas and the breast perimeter. Breast volume was measured according to the validated protocol described by Veitch et al25 and Yip et al.22 Figure 2 illustrates the application of landmarks on a large-breasted woman to section the breast from the torso for volume analysis.

Water displacement measurements based on Archimedes’ principle was performed on each breast to determine the volume, as described by Schultz et al.12 A large calibrated container was filled with warm water and the patient asked to individually immerse each breast into the container, using a specially modified surgical bed and ensuring that the superior and lateral edges of the breast were at water level (Figure 1). Individual breast volumes were then determined by the amount of water displaced.

Figure 1: Water displacement technique

Figure 2: Volumetric analysis of the breast with Cyberware WBX 3D scan and CySlice software

A=Photograph showing landmark placement following determination of breast borders by palpation
B=Scan image with outline of breast and torso boundary C=CySlice image to section the breast portion of the scan from the torso

Statistical analysis

Data was entered into a spreadsheet and statistical analysis performed using SPSS v23.0 statistical software (IBM Corp, Armonk, New York). Pearson coefficients were calculated to assess the linear association between the two measurement techniques.

The strength of any correlation between the two techniques was then determined with a Bland-Altman analysis.26 This analysis plots the mean breast volume measurements for both techniques against their differences. If the two methods are comparable and in agreement, then the differences should be small and the mean of the differences should be close to zero. The 95% limits of agreement were formed using the mean difference in volumes ±1.96 standard deviation (SD) of the differences in volumes. Linear regression analysis was performed to assess proportional bias between the two measurement techniques.

Results

A total of 322 breasts were measured in the study using both 3D laser scanner and water displacement techniques—194 breasts (97 women) were measured preoperatively and 128 breasts (64 women) were measured postoperatively. The mean breast volume according to 3D laser scan was 1440 millilitres (SD = 588 millilitres) and using water displacement was 1419 millilitres (SD = 811 millilitres).

In comparing the breast volumes obtained from 3D laser scanning and water displacement, a Pearson correlation demonstrated a strong positive linear association between the two methods (r = 0.89, p < 0.001) (Figure 3).

Figure 3: Pearson correlation of breast volume measurement—water displacement
mean volume versus 3D laser scan mean volume

There is a linear association between the two methods

Figure 4 shows the results of the Bland-Altman plot showing the differences between the two techniques plotted against the averages of the two techniques. A mean difference of -21.7 millilitres with a standard deviation of 399 millilitres was found between the two techniques. While most values were within the 95 percent limits of agreement (a lower limit of -804 millilitres and an upper limit of 760 millilitres), the majority of the data points are widely spread across this range and do not lie close to zero, which would have been the case if there was good agreement between the measured values.

Linear regression analysis confirmed a degree of proportional bias, meaning that one method gave values that were consistently higher or lower than those of the other method. In this instance, the breast volumes obtained from water displacement tended to be larger than those from the 3D scanner, and the difference between the two methods increased as the mean volume increased. This is evident from the plot where there was a discernable trend for more points to be above the line of mean difference in volume as the mean volume increased (Figure 4).

Figure 4: Bland-Altman plot showing mean breast volume versus difference in mean
breast volume as measured by water displacement and 3D scanning techniques

While most values lie within the 95% limits of agreement on the plot (±1.96 x SD), there is a large
difference in measurement of breast volume using the two methods

The statistical analyses therefore suggest that despite a strong linear correlation between the two methods of breast volume measurement, the measures have low agreement on actual values. The water displacement values were consistently larger than the 3D scan values. The greatest agreement in breast volume measurement between the two methods (within 10 percent) was seen for volumes between 1000 and 2000 millilitres (Figure 5). The findings of this study are in contrast to previous studies conducted in our unit that demonstrated equivalence between 3D laser scanner and direct water displacement of mastectomy specimens, and also between the 3D laser scanner and MRI.

Figure 5: The level of agreement between the two methods

Solid green line indicates volumes within 10% agreement and orange dashed line indicates volumes within
20% agreement. Shaded area highlights the range of breast volumes with the highest level of agreement.

Figure 6 demonstrates the breast volumes measured using water displacement and 3D laser scanning in a single patient both preoperatively and postoperatively at 12 months. The amount of breast tissue removed from each breast was 1444 grams, totalling 2888 grams. The patient’s body weight remained relatively stable during this period—127 kilograms prior to surgery and 125.3 kilograms at the postoperative time point.

Figure 6. Breast volumes measured using water displacement and 3D laser scanning
in a single patient both preoperatively and postoperatively at 12 months

Discussion

A large number of techniques for the objective measurement of breast volume have been described in the literature. However, these techniques often vary in accuracy and reliability and some are limited in their application to large-breasted patients. Many studies specifically on women with breast hypertrophy aim to develop a formula for estimating breast reduction weight in reduction mammoplasty rather than the assessment of breast volumes both before and after surgery.27-32 This is most likely driven by restrictions placed on breast reduction surgery by healthcare providers or insurance companies which often require surgeons to predict a minimum weight of resected breast tissue. Although a formula to predict tissue resection weight is clinically useful, there remains a need for the accurate determination of absolute breast volumes both in planning breast surgery and in assessing outcomes following surgery.

The use of 3D scanning in the specialty of plastic and reconstructive surgery has increased in recent times. It provides a valuable and non-invasive assessment of the breast for the surgeon, describing the volume, shape, contour, projection and symmetry. This information can not only be useful in the preoperative phase but in the assessment of postoperative outcomes and symmetry. The accurate and objective measurement of breast volume is particularly important in the complex field of breast reconstruction, such as in assessing outcomes following autologous fat grafting techniques16 and preoperatively for unilateral reduction or augmentation for asymmetry.33,34 Three-dimensional scanning technology has also been employed in other areas of plastic surgery including facial symmetrisation 35,36 and for the preoperative planning of nasal surgery.37

The Cyberware WBX laser scanner and CySlice software used in this study have previously been validated against water displacement in measuring mastectomy specimens with a strong correlation demonstrated.22 The same 3D laser scanner and protocol were also found to be strongly equivalent to non-contrast MRI for the assessment of breast volume.16

The most accurate metric of true breast tissue volume is water displacement of mastectomy specimens, however, this is not applicable other than in post-mastectomy reconstruction scenarios.20,22,38 Additionally, volume measurement of the intact breast has proven to be more variable due to the challenge of correctly differentiating breast tissue from the chest wall.

This study aimed to determine whether water displacement is a valid lower-cost method of breast volume measurement in women with breast hypertrophy compared to 3D laser scanning. Our measurement of 322 breasts by both 3D laser scan and water displacement showed a close association between the two techniques but the water displacement method had a clear tendency to estimate a larger volume than the 3D scan. A possible explanation for the difference between techniques is that measurement using water displacement requires the patient to be in a prone position, which may change the form of the breast and add axillary or abdominal fat to the breast measurement. In contrast, patients are in the upright position for measurement with the 3D scanner.

The water displacement technique is also highly dependent on the positioning of the patient. Alternative devices for the water displacement technique have been described to address the disadvantages of the immersion technique used in this study.39 Our study supports previous findings that, although water displacement is easily accessible and a low-cost option, the technique has relatively low accuracy for measurement of the intact breast due to poor reproducibility.18,40 Three-dimensional scanning and MRI display the highest level of accuracy and reliability and provide a comprehensive assessment of volume and shape for breasts in situ. However, these techniques can be cost-prohibitive. Linear anthropometric measurements are less accurate, however, the costs are negligible, they are fast and they have been shown to provide a reliable estimate of breast volume in the clinical setting.14,15,19,41

Another source of difference is that 3D scanning technology allows for a curved cut plane at the back of the breast where it can digitally follow the curved chest wall, while water displacement has by definition a flat cut plane at the water level. The manual positioning of landmarks prior to 3D scanning using palpation to accurately identify the margins of the breast base, and relying on these points later in the scan image, helps to control for potential location error.

Key advantages of this study are its large sample size and that the data was collected prospectively. One limitation of this study is that not all participants had breast volume measurements taken at both the preoperative and 12 months postoperative times. It was decided that additional measurement was too onerous for some patients given the commitments associated with voluntarily participating in the study. As a result, both of these techniques were performed on discrete subsets of patients at each time point.

Future extensions to this work would be to incorporate the results of comprehensive anthropometric body shape assessment that was also conducted pre- and postoperatively as part of this study. Analysis of patient-reported outcome measures data collected during the study would also further enhance our understanding of surgery outcomes and of the overall satisfaction in women with reduction in breast volume through surgery.

Conclusion

This study found a strong linear association between the measurement of breast volumes using water displacement and 3D laser scanning techniques in women with breast hypertrophy but the methods have low agreement on actual values. While water displacement may be more convenient and accessible in clinical practice, 3D scanning remains preferable as it has been proven to be a more accurate and reliable technique for the determination of intact breast volume.

Acknowledgements

The authors would like to acknowledge and thank the following people: members of the AFESA research group; Andrea Smallman, nurse practitioner; David Summerhayes and Lynton Emerson for clinical photography; Chris Leigh for the water displacement apparatus; Phil Dench from Headus Australia for the design and volume calculation software; Kathleen Robinette for assistance in scanner selection; Professor David Watson and Professor Julie Ratcliffe for PhD supervision; the Australian Government Research Training Program Scholarship for providing a PhD scholarship at Flinders University.

Disclosure

The authors have no financial or commercial conflicts of interest to disclose.

References

  1. Kerrigan CL, Collins ED, Striplin D, Kim HM, Wilkins E, Cunningham B, Lowery J. The health burden of breast hypertrophy. Plast Reconstr Surg. 2001;108(6):1591–99.
  2. Singh KA, Losken A. Additional benefits of reduction mammaplasty: a systematic review of the literature. Plast Reconstr Surg. 2012;129(3):562–70.
  3. Collins ED, Kerrigan CL, Kim M, Lowery JC, Striplin DT, Cunningham B, Wilkins EG. The effectiveness of surgical and nonsurgical interventions in relieving the symptoms of macromastia. Plast Reconstr Surg. 2002;109(5):1556–66.
  4. Mello AA, Domingos NA, Miyazaki MC. Improvement in quality of life and self-esteem after breast reduction surgery. Aesthet Plast Surg. 2010;34(1):59–64.
  5. Blomqvist L, Eriksson A, Brandberg Y. Reduction mammaplasty provides long-term improvement in health status and quality of life. Plast Reconstr Surg. 2000;106(5):991–97.
  6. Krieger LM, Lesavoy MA. Managed care’s methods for determining coverage of plastic surgery procedures: the example of reduction mammaplasty. Plast Reconstr Surg. 2001;107(5):1234–40.
  7. Klassen A, Fitzpatrick R, Jenkinson C, Goodacre T. Should breast reduction surgery be rationed? A comparison of the health status of patients before and after treatment: postal questionnaire survey. Brit Med J. 1996;313(7055):454–57.
  8. Koltz PF, Frey JD, Langstein HN. Insurance coverage and reduction mammaplasty: a systematic review of current health care policies. Plast Reconstr Surg. 2013;132(4):692e–3e.
  9. O’Blenes CAE, Delbridge CL, Miller BJ, Pantelis A, Morris SF. Prospective study of outcomes after reduction mammaplasty: Long-term follow-up. Plast Reconstr Surg. 2006;117(2):351–58.
  10. Seitchik MW. Reduction mammaplasty: criteria for insurance coverage. Plast Reconstr Surg. 1995;95(6):1029–032.
  11. Nguyen JT, Wheatley MJ, Schnur PL, Nguyen TA, Winn SR. Reduction mammaplasty: a review of managed care medical policy coverage criteria. Plast Reconstr Surg. 2008;121(4):1092–100.
  12. Schultz RC, Dolezal RF, Nolan J. Further applications of Archimedes’ principle in the correction of asymmetrical breasts. Ann Plast Surg. 1986;16(2):98–101.
  13. Smith DJ, J., Palin WE, J., Katch VL, Bennett JE. Breast volume and anthropomorphic measurements: normal values. Plast Reconstr Surg. 1986;78(3):331–35.
  14. Qiao Q, Zhou G, Ling Y. Breast volume measurement in young Chinese women and clinical applications. Aesthet Plast Surg. 1997;21(5):362–68.
  15. Longo B, Farcomeni A, Ferri G, Campanale A, Sorotos M, Santanelli F. The BREAST-V: A unifying predictive formula for volume assessment in small, medium, and large breasts. Plast Reconstr Surg. 2013;132(1):1e–7e.
  16. Howes BH, Watson DI, Fosh B, Yip JM, Kleinig P, Dean NR. Magnetic resonance imaging versus 3-dimensional laser scanning for breast volume assessment after breast reconstruction. Ann Plast Surg. 2017;78(4):455–59.
  17. Lee WY, Kim MJ, Lew DH, Song SY, Lee DW. Three-dimensional surface imaging is an effective tool for measuring breast volume: a validation study. Arch Plast Surg. 2016;43(5):430–37.
  18. Bulstrode N, Bellamy E, Shrotria S. Breast volume assessment: comparing five different techniques. Breast. 2001;10(2):117–23.
  19. Sigurdson LJ, Kirkland SA. Breast volume determination in breast hypertrophy: an accurate method using two anthropomorphic measurements. Plast Reconstr Surg. 2006;118(2):313–20.
  20. Losken A, Seify H, Denson DD, Paredes AA, Jr., Carlson GW. Validating three-dimensional imaging of the breast. Ann Plast Surg. 2005;54(5):471–76; discussion 7–8.
  21. Kovacs L, Eder M, Hollweck R, Zimmermann A, Settles M, Schneider A, Endlich M, Mueller A, Schwenzer-Zimmerer K, Papadopulos NA, Biemer E. Comparison between breast volume measurement using 3D surface imaging and classical techniques. Breast. 2007;16(2):137–45.
  22. Yip JM, Mouratova N, Jeffery RM, Veitch DE, Woodman RJ, Dean NR. Accurate assessment of breast volume: a study comparing the volumetric gold standard (direct water displacement measurement of mastectomy specimen) with a 3D laser scanning technique. Ann Plast Surg. 2012;68(2):135–41.
  23. Choppin SB, Wheat JS, Gee M, Goyal A. The accuracy of breast volume measurement methods: A systematic review. Breast 2016;28:121–29.
  24. Bouman FG. Volumetric measurement of the human breast and breast tissue before and during mammaplasty. Br J Plast Surg. 1970;23(3):263–64.
  25. Veitch D, Burford K, Dench P, Dean N, Griffin P. Measurement of breast volume using body scan technology (computer-aided anthropometry). Work. 2012;41 Suppl 1:4038–045.
  26. Bland JM, Altman DG. Comparing methods of measurement: why plotting difference against standard method is misleading. Lancet. 1995;346(8982):1085–087.
  27. Eder M, Papadopulos NA, Kovacs L. Breast volume determination in breast hypertrophy. Plast Reconstr Surg. 2007;120(1):356–57.
  28. Kocak E, Carruthers KH, McMahan JD. A reliable method for the preoperative estimation of tissue to be removed during reduction mammaplasty. Plast Reconstr Surg. 2011;127(3):1059–064.
  29. Sommer NZ, Zook EG, Verhulst SJ. The prediction of breast reduction weight. Plast Reconstr Surg. 2002;109(2):506–11.
  30. Kececi Y, Sir E. Prediction of resection weight in reduction mammaplasty based on anthropometric measurements. Breast Care. 2014;9(1):41–45.
  31. Descamps MJ, Landau AG, Lazarus D, Hudson DA. A formula determining resection weights for reduction mammaplasty. Plast Reconstr Surg. 2008;121(2):397–400.
  32. JZ, Wendel JJ, Zellner EG, Hagan KF, Shack RB, Corlew DS. Association between preoperative measurements and resection weight in patients undergoing reduction mammaplasty. Ann Plast Surg. 2010;64(5):512–15.
  33. Yip JM, Watson DI, Tiggemann M, Hsia S, Smallman AE, Dean NR. Determinants of breast reconstruction outcome: how important is volume symmetry? J Plast Reconstr Aes. 2015;68(5):679–85.
  34. Liu C, Luan J, Mu L, Ji K. The role of three-dimensional scanning technique in evaluation of breast asymmetry in breast augmentation: a 100-case study. Plast Reconstr Surg. 2010;126(6):2125–132.
  35. Kovacs L, Zimmermann A, Brockmann G, Gühring M, Baurecht H, Papadopulos NA, Schwenzer-Zimmerer K, Sader R, Biemer E, Zeilhofer HF. Three-dimensional recording of the human face with a 3D laser scanner. J Plast Reconstr Aes. 2006;59(11):1193–202.
  36. Knoops PGM, Beaumont CAA, Borghi A, Rodriguez-Florez N, Breakey RWF, Rodgers W, Angullia F, Jeelani NU, Schievano S, Dunaway DJ. Comparison of three-dimensional scanner systems for craniomaxillofacial imaging. J Plast Reconstr Aes. 2017;70(4):441–49.
  37. Lekakis G, Claes P, Hamilton GS 3rd, Hellings PW. Three-dimensional surface imaging and the continuous evolution of preoperative and postoperative assessment in rhinoplasty. Facial Plast Surg. 2016;32(1):88–94.
  38. Kayar R, Civelek S, Cobanoglu M, Gungor O, Catal H, Emiroglu M. Five methods of breast volume measurement: a comparative study of measurements of specimen volume in 30 mastectomy cases. Breast Cancer Basic Clin Res. 2011;5:43–52.
  39. Tezel E, Numanoglu A. Practical do-it-yourself device for accurate volume measurement of breast. Plast Reconstr Surg. 2000;105(3):1019–023.
  40. Henseler H, Khambay BS, Bowman A, Smith J, Paul Siebert J, Oehler S, Ju X, Ayoub A, Ray AK. Investigation into accuracy and reproducibility of a 3D breast imaging system using multiple stereo cameras. J Plast Reconstr Aes. 2011;64(5):577–82.
  41. Brown N, White J, Milligan A, Risius D, Ayres B, Hedger W, Scurr J. The relationship between breast size and anthropometric characteristics. Am J Hum Biol. 2012;24(2):158–64.